RFID devices interfere with medical electronics, says report





Courtesy of EE Times Europe

LONDON — The use of radio frequency identification (RFID) in hospitals can interfere with critical care medical equipment according to a study conducted by University of Amsterdam researchers published in the Journal of the American Medical Association.

The research was done in May 2006 and one active (125 kHz) and one passive (868 MHz) RFID system were operated at about 30 centimeters (1 foot) from each of 41 pieces of medical equipment at the Academic Medical Centre, University of Amsterdam, Amsterdam, the Netherlands.

In 123 tests for electrogmagnetic interference, RFID induced 34 EMI incidents: 22 were classified as hazardous, two as significant, and 10 as light. The passive system at 868 MHz induced 26 of the incidents while the active 125-kHz RFID signal caused 8 incidents in 41 EMI tests.

One conclusion of the study is that the implementation of RFID in the critical care environment should require on-site EMI tests and updates of international standards.

Another result is that it is unlikely that RFID can be adopted quickly for use in patient or medical equipment identification or tracking.

Related articles:

Startup implants RFID in operating room

E-medicine on global trajectory

Testbed streamlines RFID development



 






 Featured Jobs
D. E. Shaw Research seeking ASIC Design in New York, NY

ON Semi seeking Sr Mixed-Signal Test Development Engr in Lower Gwynedd, PA

SanDisk seeking Sr Process Integration Engineer in Milpitas, CA

Lowes seeking ITO Systems Engineer I in San Antonio, TX

Satcon seeking Sr. Power Electronics Engineers in Linthicum, MD

More jobs on EETimesCareers
 Sponsor
 CAREER CENTER
Ready to take that job and shove it?
SEARCH JOBS:

 SPONSOR

 RECENT JOB POSTINGS
For more great jobs, career related news, features and services, please visit EETimes' Career Center.